Clinical Trial

Alcohol Monitor Validation

Study Description

Alcohol Monitor Validation

This study will validate a wrist-worn alcohol monitor (BACtrack Skyn) in both laboratory and real-life settings.

Location

Locations Selected Location

Methods

No pharmaceutical medication involved No pharmaceutical medication involved
Patients and healthy individuals accepted Patients and healthy individuals accepted

Device - Skyn Monitor Lab Session 1

Participants will be instructed to wear the Skyn wrist-worn alcohol monitor and drink 3 beers over the course of an hour and a half. BrAC will be measured every 15 minutes throughout the session.

Device - Skyn Monitor Field Test with EMA App

After Skyn Monitor Lab Session 1, participants will be instructed to wear Skyn monitor for the rest of the day and the following two weeks. Participants will also enter information on each drinking session during the two week time using a mobile-based ecological momentary assessment (EMA) to collect real-time data in daily life.

Device - Skyn Monitor Lab Session 2

Participants will be instructed to wear the Skyn wrist-worn alcohol monitor and drink 3 beers over the course of an hour and a half. BrAC will be measured every 15 minutes throughout the session. Participant will also complete device acceptability and usability survey.

Additional Information

Official Study Title

Laboratory and Field Validation of a Wrist-Worn Alcohol Monitor

Clinical Trial ID

NCT03617705

ParticipAid ID

dyP8Eb